If you’re having problems with the water pump within your Tuttnauer autoclave you need to find out what the thing is right away and fix it. No water or a water level that’s as well reduced means your machine isn’t functioning properly and won’t be completely sterilizing your gear. And that means you’re essentially away from company until you get the problem resolved.
If you’ve constantly sent your autoclave out for repair anytime there’s a difficulty then you definitely know just how expensive these fixes can be. You may have work costs, costs for parts, and don’t forget the shipping and handling charges. You might nearly purchase a new machine when all has been said and completed! If you’re having difficulty with the water in your Tuttenauer autoclave, here are some problem solving ideas that might enable you to solve the situation your self.
In the event the water is flowing within the incorrect path, check both pipes connected to the water pump. It may be which you just need to reverse them.
In the event the pump is making noise check to see if the mounting bracket is loose or damaged. You may just need to tighten it down or change it out. If it’s not the bracket the problem is most likely in the water pump alone and you’ll require a new one.
When the water pump helps to keep coming a fuse, initially search for pinched or twisted wires that might be resulting in a short. If that’s not the situation, turn the water pump away and disconnect the wires from the number 1 and two terminals in the strong state relay. Transform the water pump back on and in case the fuse blows once again the thing is with either the water water pump or perhaps the water pump capacitor and you’ll must change both.
When the pump won’t switch off then you need a voltage/ohmmeter to check the communicate. Read between TP13 and TP1. A 10 to 12 volt DC transmission is required to tell the relay to change away along with a to 1 transmission is required to tell it to change on. When the reading through is everything else it means you do have a problem with the control circuit.
In the event the transmission is proper, repeat the process at the communicate. And in case you obtain an incorrect transmission there then you’ll need to replace the Ajunc board.
If your pump won’t switch on, the very first thing you need to do is check the fuse and replace it if necessary. Following, try to find pinched or twisted wires leading from your solid state relay for the pump. If ejxawh doesn’t help, turn the ability to your autoclave off and disconnect the cables on terminals 1 and 2 around the communicate. Now, link those two wires together and transform the water pump back on. When the pump still doesn’t work then you’ll need to change it.
While your autoclave might become a ‘machine’ and it also executes a very important, life-saving function, it’s nevertheless, in the end, just a large box with a few fundamental operating parts. It’s typically easy enough to find the issue. And most of the time you can correct it yourself!